Location: /Products by Categories/Operating Room/Surgical Room/Anesthesia Machines

Narkomed Drager 2B

Narkomed Drager 2B

Product Information

The NAD Narkomed 2B is one of their longest running and most reliable model lines. The 2B has O2 and N2O standard with optional medical air and holds 3 vaporizers. An AVE ventilator with F-Style ascending bellows is a standard feature with oxygen, volume & pressure monitoring. Drawer pack configurations may vary from picture. Features and Options Overview:

  • Choose 2 of 19.1 vaporizers available: ORM standard; ORMC optional
  • Ventilator options: Pneumatic or electronic, ascending or descending bellows
  • Oxygen monitor and CO2 absorber standard
  •  AVE Electronic ventilator
  • Absorber and scavenger
  •  Oxygen analyzer
  •  Breathing pressure monitor
  •  Respiratory volume
  •  Centralized alarm and data
  • Oxygen ratio controller
  •  Pressure limit controller

Dimension Specifications

  •  Maximum Height: 68" (173cm)
  •  Maximum Width: 40" (102cm)
  •  Maximum Depth: 25" (63.5cm)
  •  Table-top Height: 34" (86cm)
  •  Table-top Area: 271 in. sq. (1,749cm2)
  •  Shelf Area: 200 in. sq. (1,290cm2)
  •  Drawer Dimensions
    • Large: 14"x21"x7" (36x52x18cm)
    • Small: 14"x17"x3" (36x43x8cm)
  •   Nominal Weight: 325 lb. (147kg)


Electrical Conductivity:
The NARKOMED 2B Anesthesia System is equipped with conductive casters. All anesthesia machines are tested for electrical conductivity between one caster as a measuring point and the table-top, shelf-top, and inspiratory valve, alternately, as the other measuring point. The maximum allowable resistance between any two points is 250,000 ohm.

Gas Delivery System Information:

Standard Gases: The Narkomed 2B is equipped with pneumatic circuitry for the delivery of oxygen (o2) and nitrous oxide (N2O). It has at least on oxygen and one nitrous oxide yoke for reserve gas cylinders with flush-type valves. The Pin Index Safety System prevents connection of a cylinder of the incorrect type. Option Gases: In addition to oxygen and nitrous oxide, the NARKOMED 2B may be equipped with up to two additional gases. The additional gas may be air, helium, (He), nitrogen (N2), or carbon dioxide (CO2). The additional gas may be supplied to the anesthesia system by means of pin-indexed cylinders and yokes, by Diameter Indexed Safety System (DISS) pipeline connections, or by both systems, if so selected.

Color Coding: Each connection, valve, gauge, and flowmeter is labeled and color-coded for the appropriate gas, as shown in the following table:

 

Gas

Label

Color Code

Air

AIR

Yellow

Carbon Dioxide

CO2

Gray

Helium

He

Brown

Nitrogen

N2

Black

Nitrous Oxide

N2O

Blue

Oxygen

O2

Green

Manufacturer: NarKomed